Introduction to Fixed Oxygen O2 Gas Detectors

  

The SA-3003 Fixed Oxygen Detector is an on-site detection instrument with concentration display and audible/visual alarm. It is primarily used to detect various volatile organic compounds in open or confined environments, and the on-site concentration values can be transmitted in real-time to remote monitoring systems.

 Fixed Oxygen O2 Gas Detector Product Features:

      DC24V power supply;

      Two output signal options: three-wire (4–20)mA and RS485;

      Utilizes electrochemical sensor with high sensitivity, good linearity, long lifespan, and stable performance;

      Optimized circuit design with built-in reverse polarity protection and current limiting protection to prevent unnecessary damage from wiring errors;

      One or two passive relay contact output signals for driving external devices such as fans and solenoid valves;

      Explosion-proof rating: Exd II CT6; protection rating: IP65;

      Housing material: die-cast aluminum + 304 stainless steel (sensor cover);

      Compatible with various gas detection alarm controllers, PLC, DCS, and other control systems for remote monitoring.

 

Fixed Oxygen O2 Gas Detector Technical Parameters:

 

 

Name

Fixed Oxygen Detection Alarm

Model

SA-3000-O2, SA-3002-O2, SA-3003-O2

Verification Regulation

JJG 365-2008 "Verification Regulation for Electrochemical Oxygen Analyzers"

Detection Principle

Electrochemical

Concentration Unit

%VOL

Operating Voltage

DC24V or AC220V

Output Signal (Optional)

Standard (4–20)mA current signal, RS485 digital signal, 1 to 2 sets of passive normally open relay signals

Explosion-Proof Rating

Ex dII CT6

Protection Rating

IP65

Wiring Method

Split-type: three-wire system; Bus-type: four-wire system

Dimensions

178*139.5*106mm

Interface Type

1/2 inch internal thread

Instrument Parameters

Conventional

Unconventional

Range

(0–30)%VOL

 

Resolution

0.2%VOL

 

Operating Temperature

(-20–50)℃

 

Operating Humidity

(0–95)%RH (non-condensing)

 

Operating Pressure

(90–110)kPa

 

Response Time T90

≤15s

 

Indication Error

±3%FS

 

Repeatability

≤1%

 

Zero Drift

±1%FS

 

Span Drift

±1%FS

 

 

 

Application Fields:

Petrochemical industry, industrial production, smelting and forging, power industry, coal mining, tunnel engineering, environmental monitoring, sewage treatment,

Paper manufacturing, warehousing and logistics, brewing and fermentation, fire protection, gas industry, waste treatment plants, school laboratories, research centers

Industry Category Measurement-Analysis-Instruments
Product Category
Brand: 常州顺安
Spec: SA-3003
Stock:
Origin: China / Jiangsu / Wuxishi
About Toocle.com - Partner Programme - Old Version
Copyright © Toocle.com. All Rights Reserved.
(浙)-经营性-2023-0192